Selecting Hold Conditions for Measurement Values

The hold function allows holding (maintaining) an arbitrary value, such as the maximum or minimum
values, among the measurement values obtained from the start to the end of the measurement, and
outputting the value after the measurement is completed.
Example 1: Holding the peak value (used in combination with hold output)
Measurement
value
Start of measurement
How to set the timing to trigger hold measurement
Example 2: Holding the peak value for every three measurements (used in combination with the logging
function)
1
Measurement
Measurement
value
How to set the logging function
It is possible to hold the following five types of values. The item to be held can be specified for each
output (OUT0 to 7).
Item
Sampling
The value at the start of the measurement is held.
Peak
The highest value measured during the measurement is held.
Bottom
The lowest value measured during the measurement is held.
Peak-Peak
The difference between the highest and lowest values measured during
the measurement is held.
Average
The average between the highest and lowest values measured during
the measurement is held.
None
Hold measurement is not performed.
Setting Measurement Conditions (Expert Menu)
Hold the peak value
2
3
1
2
Peak value
Accumulate and output
the held value.
p.126
Meaning
Output the held value
at the end of the measurement.
End of measurement
p.118
3
1
2
Peak value
Peak value
Accumulate and output
Accumulate and output
the held value.
SECTION 3
3
1
the held value.
Z550
Operation Manual
111