TREND (or T)
Function......Obtains the measurement values from the start to the end of the measurement
under the Trigger setting.
(Cannot be performed when the setting of [Hold&Logging] is other than [Free].)
TREND (or T) <Output> <Output coefficient>
Input
• Specify an output between 0 and 7.
Output #
0
1
2
3
4
5
6
7
• Specify an output coefficient between 1 and 1000.
The output coefficient indicates the intervals between measurements that are
output.
The number of data items that will be output can be calculated using the formula
below.
In the case of hold output
Measurement period ÷ Measurement time = Number of measurements
Number of measurements ÷ Measurement coefficient = Number of data items
output (The number is rounded up to the nearest integer.)
In the case of logging output
Measurement period ÷ Measurement time ÷ Sampling interval = Number of
measurements
Number of measurements ÷ Measurement coefficient = Number of data items
output (The number is rounded up to the nearest integer.)
Scn 0
Run
(from start to end of measurement)
Meaning
OUT0
OUT1
OUT2
OUT3
OUT4
OUT5
OUT6
OUT7
XX.XXms
Measurement time
Measurement period
Communication with External Devices
Delimiter
Measurement time
Measurement result
SECTION 6
Z550
215
Operation Manual