The measurement screen for point 1 appears.
5.
Position the workpiece correctly and press the ENT
Key.
The screen for selecting the position of point 1 appears.
6.
Set the value to 0.00000 and press the ENT Key.
The screen for selecting the position of point 2 appears.
7.
Set the value to 5.00000 and press the ENT Key.
A confirmation message appears.
8.
Select [Execute].
The setting will be registered and the screen display returns to
Setting Measurement Conditions (Expert Menu)
System will be calibrated.
Execute