A confirmation message appears.
9.
Select [Execute].
The setting will be registered and the screen display returns to
Compensate for the Measured Level Difference and Width
To compensate for the error, perform the measurement once.
Applicable measurement items: Step: 2 pts, Step: 3 pts, Width
Example: In case of [Step: 2 pts]
Difference: 5 mm
Compensate for a measurement value of 4.95 mm, such that it is registered as 5 mm.
Point 2
To compensate for this error, perform the measurement once.
Measure with the
workpiece in position.
Setting Measurement Conditions (Expert Menu)
5mm
4.95mm
Calibrate to 5 mm.
Point 1
Input the value
of point 1.
0mm
System will be calibrated.
Execute
3
.
Measurement value
Compensate for
this error.
Level difference
workpiece
5mm
Input the value
of point 2.
5mm