Input Format - Omron Z550 Operation Manual

Multi-dimensional sensor
Hide thumbs Also See for Z550:
Table of Contents

Advertisement

SECTION 6
Communication with External Devices

Input Format

The following types of input are available from the terminal blocks in the Run mode.
Terminal name
RESET
TRIGGER
LDOFF
DI0
DI1
DI2
DI3
DI4 to DI7
* Scene switching
Scene
Scene 0
Scene 1
Scene 14
Scene 15
0: Off status
1: On status
The time required to turn on an input terminal is "9.96 ms x 2" or more (excluding scene switching).
Z550
190
Operation Manual
The measurement values are cleared and the Controller is restarted.
Settings that have not been saved are cleared, and the setting data saved
in the flash memory is loaded.
Triggers measurement.
The laser diode (LD) is turned off.
Force-zero is performed for OUT0.
To cancel the force-zero function, turn this terminal on for at least 1
second.
When the force-zero function is executed, the zero value is determined
using the latest measurement value even if the hold output or logging
output is specified.
Force-zero is performed for all outputs.
To cancel the force-zero function, turn this terminal on for at least 1
second.
When the force-zero function is executed, the zero value is determined
using the latest measurement value even if the hold output or logging
output is specified.
(Not used)
Used in combination to switch scenes.*
The scene switching time is approximately 2 seconds.
Change the setting of [Environment/Scene] of the System mode to
[Terminals] in advance.
p.177
DI7
DI6
DI5
0
0
0
0
0
0
1
1
1
1
1
1
Function
DI4
0
1
0
1