SECTION 3
Setting Measurement Conditions (Expert Menu)
When Selecting [Fixed]
Position a reference workpiece and select an appropriate sensitivity level while checking the light
reception level.
Setting range
LV0 (laser turned off)
LV31
1.
Select [Image/Sensitivity].
The selections for sensitivity adjustment appear.
2.
Select [Fixed].
The screen for setting the sensitivity level appears.
3.
Select the sensitivity level using the Up/Down keys
while monitoring the image on the monitor screen.
Select
brightness and light reception level.
• 120 or less: Insufficient light density
• 255 or more: Excessive light density
4.
Select the setting by pressing the ENT Key.
The setting will be registered and the screen display returns to
Z550
52
Operation Manual
Sensitivity
-
LV1
Low
·
·
·
High
the
level
while
checking
Color of
workpiece
-
Bright
Dark
Scn 0
Set
Image
Meas method
Detect set
Region
Meas set
Sensitivity
Extra set
Details
Meas
mode
Sensitivity level
NEAR
LV 15
the
line
FAR
DARK
Line Brightness
1
.
Auto
Fixed
Multi
Received level
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
[
]
LV
[
]
PEAK
BRIGHT
Line beam image