Pc Programming Station (Pcps); Programmable Parameters - Otto Bock C2000 Service Instructions Manual

Table of Contents

Advertisement

c) Changing values
Navigate through the menu tree by moving the blinking cursor using the four-directional button.
The respective sublevel can be opened by using the right button in each node until you reach
the parameter value level. Parameters can be increased or decreased by pressing the plus or
minus symbol on the rocker button. By pressing the rocker button briefly, the values are changed
by small increments; keeping the rocker button depressed for a longer time changes the values
by larger increments.
The following list shows the programmable wheelchair parameters.
The monitor values, list of individual modules errors, sub-item Information as well as handheld
programming device settings are not listed here individually. However, they are easily viewed
on the handheld programming device thanks to the uniform menus.
INfORMATION
Ensure that the freely accessible parameter for the standard LCD contrast (default value 130) is not
changed, because the value quickly leaves the defined range and the font can no longer be distin-
guished from the background. In that case the handheld programming device must be reset using a
PC programme.

8.4.2 PC Programming Station (PCPS)

The enAble50 control unit can also be programmed with the PCPS 1314 programming station.
In addition to the programming software, an interface is required to connect the computer and
control unit. An adapter is required to programme the enAble50 wheelchair control unit. Article
no. 493U32=SK185 includes the programming software, interface and adapter cable.
INfORMATION
For more information on using the software, please consult the enAble50 instructions for use.
8.5  Programmable Parameters
WARNING
Risk of accidents and injury due to incorrect configuration settings. Modified parameter settings in
the configuration can lead to changes in driving characteristics. In particular, changes to the speed,
acceleration, braking or joystick settings can lead to unexpected and therefore uncontrollable operat-
ing performance with a risk of accidents.
Always test the driving characteristics of the power wheelchair after configuration / programming is
complete.
Programming must only be completed by authorised personnel. Neither Otto Bock nor the control unit
manufacturer are liable for damages (especially in combination with special controls) caused by pro-
gramming that was not properly / professionally adapted to the abilities of the wheelchair user.
The enAble50 wheelchair control unit includes a number of parameters that can be programmed
using the handheld programming device or the PCPS 1314 (see Section 8.4). These program-
mable parameters allow the driving characteristics and performance of the vehicle to be adapted
to special user requirements. Information about the use of the handheld programming device
can be found in Section 8.4.1.