MTU Input Queue
The MTU Input Queue is used to load MTUs. A maximum of 125 tests worth of tubes (25
MTUs) can be loaded onto the instrument.
Distributor
The Distributor transfers the MTUs to the various components located within the system.
Its transfer mechanism is based on a hook and rail system that pushes and pulls the
MTUs between the instrument modules.
AMP Load Station
The AMP Load Station is used for the addition of Amplification Reagent and Enzyme
Reagent into the MTUs during assay processing. This heated load station features an
orbital mixer that ensures proper mixing following reagent addition.
HPA Load Station
The HPA Load Station is used for the addition of Probe Reagent and Selection Reagent
into the MTUs during assay processing. This heated load station features an orbital mixer
that ensures proper mixing following reagent addition.
Magnetic Wash Module
The instrument contains two (2) Magnetic Wash Stations. The Magnetic Wash Stations
separate the bound target nucleic acid from potential inhibitors and cell debris. Each
wash station consists of fluid injectors that dispense Wash Solution into the MTU tubes,
aspirators that draw the supernatant from the tubes, and a magnet carriage that separates
the target nucleic acid/magnetic particle complex from the test solution. The wash stations
also dispense oil into the MTUs following the wash cycle.
Sample Mix Station
The Sample Mix Station is used for the addition of sample and working Target Capture
Reagent (wTCR) to the MTUs. It features an orbital mixer that ensures proper mixing of
the sample and wTCR.
Luminometer
The Luminometer measures the chemiluminescent reaction triggered by the addition of
Auto Detect fluids to each tube of the MTU. The luminescence is measured with a
photomultiplier tube (PMT).
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
Overview
38