Table 116. Assay Processing Flags - Hologic Panther System Operator's Manual

Table of Contents

Advertisement

Table 116. Assay Processing Flags (Continued)
Code
Description
FDF3
Auto Detect 1 pump
dispense failure, which
may be due to a
hardware or fluid
connection problem.
FDF4
Auto Detect 2 pump
dispense failure, which
may be due to a
hardware or fluid
connection problem.
FL1
Fluorometer internal
error.
FL3
Fluorometer internal
error.
FLSOR
Fluorometer signal out
of range.
FW#
Firmware error detected.
A number with a
corresponding device
will follow the flag.
HB
High background
reading in the
luminometer prior to
Auto Detect dispense,
which may be due to a
hardware or reagent
preparation problem.
IDF
Incubator door failure.
IFF (1-3)
Incubator 1-3 fan failure.
IHF (1-3)
Incubator 1-3 heater-foil
failure.
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
Appendix B. Processing Flags
Outcome
Sample is invalid.
Check all Auto Detect 1 connections. Contact
Hologic Technical Support.
Sample is invalid.
Check all Auto Detect 2 connections. Contact
Hologic Technical Support.
MTU is invalid.
Contact Hologic Technical Support.
MTU is invalid.
Contact Hologic Technical Support.
MTU is invalid.
Contact Hologic Technical Support.
Contact Hologic Technical Support.
Sample is invalid.
Contact Hologic Technical Support.
Contact Hologic Technical Support.
Contact Hologic Technical Support.
Contact Hologic Technical Support.
274