4. Select the row of the assay type to be edited and select Result Types.
The Result Type Selection window opens.
5. To add filter criteria, select Add.
6. Select the result type to hold from being automatically sent to the LIS upon test
completion for the assay type selected.
7. If adding additional filter criteria, repeat steps 5 and 6.
8. Select the X next to the result type(s) that no longer need to be held.
9. Select OK when finished editing the result holding criteria for the selected assay type.
10. The Result Type Selection window closes and the Select Assay to hold from LIS
Automatic Sending window displays the updated filter criteria for the assay type
chosen.
11. Repeat steps 4–9 for any additional assay types, as necessary.
12. When finished with updating the filter criteria for all assay types, from the Select Assay
to hold from LIS Automatic Sending window, select OK.
13. From the LIS Configuration screen, select Save after altering LIS configuration.

LIS Messages

To view messages sent to and received from the LIS:
1. From the Task screen, select Perform Maintenance.
The Maintenance Activities screen opens.
2. From the Maintenance Activities screen, select LIS Messages.
The LIS Messages window opens.
Figure 111. LIS Messages Window with ASTM 1394 Selected (Example)
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
System Administration
239