Levey-Jennings Report; Figure 88. Exceptions Report (Example) - Hologic Panther System Operator's Manual

Table of Contents

Advertisement

Levey-Jennings Report

Select Levey-Jennings Report to generate a Levey-Jennings Report based upon the
four (4) filter categories:
Assay
l
Completion Date
l
Quality Control Type
l
Display multiple master lots on same chart
l
All four (4) filter categories must be defined in order to generate the report. Make a
selection in the categories of choice and check the cells as appropriate to generate a
report that contains all data that match the filter specifications. The mean RLU and RLU
standard deviation statistics on the chart use the data from the selected controls or
calibrators across the selected time range unless viewing an EQC with a configured
target mean and standard deviation. Once the filters have been updated, select View,
Print, or Export as desired to access the Levey-Jennings Report.
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)

Figure 88. Exceptions Report (Example)

System Operation
173