Enable Power Save Mode; Time To End Power Save Mode; Ups Management - Hologic Panther System Operator's Manual

Table of Contents

Advertisement

Enable Power Save Mode

To configure the system to transition into the Power Save state:
1. From the expanded Power Save Management menu on the System Configuration
screen, select the Enable Power Save Mode field.
The menu expands to show the Minutes of inactivity before entering Power Save
Mode after Setup or Ready state field.
2. Select the desired number of minutes of inactivity before the system transitions into
the Power Save state from the Ready or Setup states.

Time to End Power Save Mode

To configure the system to automatically transition out of the Power Save state at a set
time:
1. From the expanded Power Save Management menu on the System Configuration
screen, select the Time to End Power Save field.
A window opens.
2. From the clock on the window, select the desired time for the system to exit the Power
Save state.
3. Select OK when finished.
4. Check the boxes for the days of the week that the system should automatically
transition out of the Power Save state.
5. From the clock on the window, select the desired time for the system to exit the Power
Save state.
6. Select OK when finished.
7. Check the boxes for the days of the week that the system should automatically
transition out of the Power Save state.
If the system is in the Power Save state at the selected day and time, the system will
transition out of the Power Save state at that time. It may take up to 33 minutes for
system temperatures to lock after exiting the state. All temperature-monitored
modules need to reach target temperatures before the system can enter the Ready or
Assay Processing states.

UPS Management

A UPS can be connected to the Panther System to provide back-up power to the system
for a set period of time during interruption of the main power supply. The UPS also
protects the system from potentially harmful incoming power surges.
When the system is configured with a UPS, manual shut down of the computer is not
required during a power outage. If power loss occurs, the instrument is taken into a safe
state and the computer is powered down after a short delay. If this occurs, manually turn
off the instrument.
Use the UPS Management expansion menu to configure the UPS settings as needed. 
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
System Administration
234