Appendix B. Processing Flags

The tables in this appendix contain descriptions of the codes that can appear on the
results if an error occurs during assay or results processing.

Assay Processing Flags

Table 116. Assay Processing Flags
Code
Description
APV
The internal Assay
Processes Verifier
detected a problem with
the assay process.
ARR
Assay Reagent
pipetting operation
aborted.
CFM
User defined leading
control set failed due to
a hardware error.
CL1
Calibrator invalidated
due to failure of worklist
calibrator set.
CLT
A clot was detected in
the sample tube during
aspiration.
CPF
Cap piercing failure,
which may be due to a
hardware or penetrable
cap problem.
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
Appendix B. Processing Flags
Outcome
Sample is invalid. Review must invalidate the
result.
Review additional assay processing codes to
determine error source. Check all fluid
connections. Contact Hologic Technical
Support.
Sample is invalid.
Reagent rack currently pipetted was
removed/ Reagent pipettor operation aborted.
Sample must be retested.
Sample is invalid.
Contact Hologic Technical Support.
Calibrator is invalid.
Address the cause of failure for invalid
calibrators in the worklist. Contact Hologic
Technical Support.
Sample is not tested. Sample results marked
as Invalid.
Sample must be retested. Contact Hologic
Technical Support.
Sample is not tested. Sample results marked
as Invalid.
Examine the cap for defects. Contact Hologic
Technical Support.
272