Results Processing Flags

Table 117. Results Processing Flags

Code
Description
a
Invalid kinetic profile.
The result calculation
for the flasher analyte
curve fit has failed.
af
Invalid kinetic profile.
The post-kinetic
adjustment value for the
flasher analyte is out of
range.
afc
The post-kinetic
adjustment value for the
flasher is greater than
the maximum cutoff
value for the assay.
ag
Invalid kinetic profile.
The post-kinetic
adjustment value for the
glower analyte RLU is
out of range for the
calibrator or control.
agc
The post-kinetic
adjustment value for the
glower analyte is
greater than the
maximum cutoff value
for the assay.
arl
Invalid kinetic profile.
The flasher curve R
squared value is less
than the minimum cutoff
value for the assay.
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
Appendix B. Processing Flags
Outcome
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
283