Figure 64. Add Test Orders Window (Example) - Hologic Panther System Operator's Manual

Table of Contents

Advertisement

4. Select Add.
The Test Order Assays window opens with a display of assays that are
available for the chosen specimen type.
5. Select the desired assay row(s) to add test orders to the selected specimen.
6. If selecting an assay with multiple analytes, select the analytes to be tested,
if necessary, by selecting Select Analytes. If not changing the analyte
selection, skip to Step 9.
7. Add or delete analytes as appropriate by selecting Add or X for the analytes
available on the screen.
8. Select OK to save the analyte setting and return to the Test Order Assays
window.
9. Select OK on the Test Order Assays window when finished selecting the test
order assay type(s).
10. Select the Replicates field to choose the desired number of replicates for the
additional test order.
A number keypad opens.
11. Type in the desired number of replicates to be tested with the chosen assay
type and select OK when finished.
The Replicates field updates with the desired replicate number.
12. Select Accept to accept the chosen test order(s) and replicate count.
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)

Figure 64. Add Test Orders Window (Example)

Note—Use the Select Analytes functionality to select only a single
analyte test order from assays with multiple targets, if necessary.
Remove the total RLU for the analyte that is not selected
(suppressed) by enabling the Enable RLU value suppression for
Aptima Combo 2 configuration (see Results Management on page
227).
System Operation
124