Add A New Master Lot; Figure 53. New/Activate Master Log Screen (Example) - Hologic Panther System Operator's Manual

Table of Contents

Advertisement

kit. The master lot barcode can be found on the Master Lot Barcode Sheet that is
included with each assay reagent kit. Register the master lot barcode to ensure that
only the reagents belonging to a specific master lot can be run together on the
system.
Each master lot barcode only needs to be entered once into the system. Inactivate
master lots as needed to prevent the system from recognizing the assay reagent kit.
Inactivated master lot information remains in the database, but cannot be used to run
assays.

Add a New Master Lot

To add a new Master Lot to the database:
1. Select the Load Assay Reagents icon on the Tasks screen.
The Panther Assay Reagent Bay screen opens.
2. At the bottom of the Panther Assay Reagent Bay screen, select
New/Activate Lot.
A New/Activate Master Lot window opens.
3. Using the handheld barcode scanner, scan the Master Lot barcode on the
Master Lot Barcode Sheet.
The Lot Number of the new Master Lot populates the Master Lot Barcode
text box in the New/Activate Master Lot window. The New/Activate Master
Lot window expands if the Master Lot Barcode is new to the system.
4. Scan the barcodes on the Master Lot Barcode Sheet for each component
belonging to the Master Lot. The barcodes can be scanned in any order.
As each barcode is scanned, the barcode populates the appropriate text
box in the New/Activate Master Lot window.
5. Select Accept.
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)

Figure 53. New/Activate Master Log Screen (Example)

System Operation
107