Clean The Sample Shield; Table 93. Mag Wash Cleaning Inventory - Hologic Panther System Operator's Manual

Table of Contents

Advertisement

This maintenance task requires the following inventory:

Table 93. Mag Wash Cleaning Inventory

Consumable
MTUs
Tips
Waste
Panther Universal Fluid Kit
A or B
Mag Wash
Cleaning Solution
To perform the Mag Wash Clean maintenance task:
1. Load a new bottle of Mag Wash Cleaning Solution in the cleaning solution slot
according to the Replace Mag Wash Cleaning Solution instructions, if needed (see
Replace Mag Wash Cleaning Solution on page 209).
2. Prior to the time that the task will be performed, ensure that a sufficient inventory of the
required materials will be available when this maintenance task will begin.
3. Initiate the maintenance task via one of the following two (2) methods:
Schedule a recurring date and time for this maintenance task to be automatically
l
performed following instructions in the Schedule an Automated Maintenance
Task section (see Schedule an Automated Maintenance Task on page 193).
Perform this task manually as follows:
l
a. Ensure that the instrument is in the Setup or Ready state.
b. Select the Mag Wash Clean task on the Maintenance Activities screen.
c. Select Start.
The Run Maintenance Task screen for the Mag Wash Cleaning task opens.
d. Select Start.
The Run Maintenance Task screen for the Mag Wash Cleaning task
updates with an estimated completion time.
e. Once the maintenance task has completed, select Done.

Clean the Sample Shield

Clean the Sample Shield weekly in order to properly maintain a clean environment within
the system. Perform this task when the instrument is in the Setup or Ready states.
Complete the procedure as part of Monthly Maintenance and additionally as needed.
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
Minimum Quantity
10 Tests
2 Tips
20 Tests
15 Tests
25.5 mL
Maintenance
201