System Capacities And Throughput; Table 12. Panther System Dead Space Volume Requirements; Table 13. System Capacities And Throughput - Hologic Panther System Operator's Manual

Table of Contents

Advertisement

μL, plus the dead space volume of 200 μL. See the appropriate Aptima package
insert for more information.

Table 12. Panther System Dead Space Volume Requirements

Tube/Specimen Type
Aptima Specimen Transfer Tube
Aptima Urine Specimen Transport
Tube – Urine specimens
Swab Specimen Transport Tube –
Unisex swab specimens
Swab Specimen Transport Tube –
Vaginal swab specimens
Cervical Specimen Collection and
Transport specimens
Multitest Swab Transport Tube
Primary Collection Tube – Plasma or
serum
Specimen Aliquot Tube (SAT) –
Plasma or serum
*Dead space volume differs depending on assay. Refer to the appropriate assay
package insert for dead space volume requirements.
**1.2 mL above the pellet.

System Capacities and Throughput

Table 13. System Capacities and Throughput

Assay Reagent Kits
Universal Fluids
Sample Racks
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
Specifications and Requirements
Dead Space Volume Required
1300 μL
1300 μL or 700 µL*
1300 μL
1300 μL
1300 μL
1300 μL
700 μL**
200 μL
Up to 4 assay reagent kits at a time
Fluid Kit A and B – 2,000 tests
Up to 15 tubes per rack, up to 8 racks (120
tubes)
23