Logoff; System Operation - Hologic Panther System Operator's Manual

Table of Contents

Advertisement

The Tasks screen opens.
If login is completed while the system is initializing or unloading MTUs, the activity field
on the Status Bar displays the system activity being performed and the time of completion.
All buttons on the Tasks screen are available once the system transitions to the Setup
state.

Logoff

To log off:
Select the Logoff button on the Menu Bar. From the Logoff/Shutdown screen, select
l
Logoff.
The Login screen opens. System tasks already in process continue when logged off.

System Operation

Warning—Do not re-use Tips, MTUs, Waste Bags, or Waste Covers. Use
only tips, MTUs, and Waste Covers that are supplied by Hologic. Use of
unapproved consumables may compromise assay or Panther System
performance.
Use the Tasks screen to manage the instrument setup process. The Task Buttons have
been arranged on the Tasks screen to guide the operator in preparing the instrument to
begin assay processing. Use the Tasks screen to load fluids, empty waste, access
maintenance tasks, prime the instrument, and load reagents and samples.
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
System Operation