Figure 58. Control/Calibrator Lot Entry Screen (Example)

3. Using the handheld barcode scanner, scan the top barcode on the Calibrator or
Control Lot Barcode Sheet.
4. Scan all of the barcodes on the Calibrator or Control Lot Barcode Sheet for the
new lot. The barcodes can be scanned in any order.
5. Select Accept when finished.
The new control or calibrator lots are registered in the database. The
Control/Calibrator Lot Entry window closes and the operator is returned to the
Sample Rack Bay screen.
6. Repeat these instructions if loading more than one lot of calibrators or controls.

Controls/Calibrators Setup

Refer to the appropriate assay package insert regarding requirements for control
and/or calibrator validity. Controls/calibrators can be loaded in any position in
the sample rack and the rack can be loaded in any lane of the Sample Bay.
Specimens will be pipetted when one of the following two (2) conditions have
been met:
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)

Figure 59. Sample Rack Bay Screen (Example)

System Operation
114