Manual Barcode Entry Mode; Enable Worklist Id Suffix; Table 101. Manual Barcode Entry Modes - Hologic Panther System Operator's Manual

Table of Contents

Advertisement

duplicate test order is loaded. When this option is enabled, the system will transition to
the Sample Rack Loading screen after a duplicate is loaded. The operator can then
accept or reject the duplicate, or all duplicates, as necessary. Select the Allow All
Repeats button to allow any specimen that has already been processed to be tested as
designated. If the duplicate is rejected, the tube will display in red on the Sample Rack
Bay screen and will not be processed. The operator can remove the specimen tube(s)
after rejection.
To automatically reject all duplicate test orders, check the Reject All Repeat Test Orders
cell. When this option is enabled, the system will transition to the Sample Rack Loading
screen after a duplicate is loaded. In this configuration the operator will not be able to
accept the duplicate. The tube will display in red on the Sample Rack Bay screen and will
not be processed. The operator can then remove the specimen tube(s) from the system.

Manual Barcode Entry Mode

To configure the Manual Barcode Entry Mode setting:
1. From the expanded Specimen Loading menu on the System Configuration screen,
select the Manual Barcode Entry Mode field.
A Specimen Loading window opens.
2. Select the desired entry mode from the list on the Specimen Loading window. The
choices on the table are defined below.

Table 101. Manual Barcode Entry Modes

Option
Description
Single
Manual entry can be completed with a single entry. The SID entered is
visible to the operator. If the ISBT128 setting is enabled, a check digit
entry is required when using this barcode entry mode.
Double
Manual entry can only be completed after the SID has been correctly
Visible
entered twice. The SID entered is visible on the screen during both
entries.
Double
Manual entry can only be completed after the SID has been correctly
Hidden
entered twice. The SID entered is not visible on the screen during both
entries.
3. Select OK when finished.

Enable Worklist ID Suffix

To differentiate samples in a Results Report by a worklist ID suffix, check the Enable
Non-Barcode Sample Loading configuration. In this scenario, the system automatically
groups together samples during loading based upon the sample default settings (see
Configuring Additional Sample Defaults on page 119).
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
System Administration
226

Advertisement

Table of Contents
loading