The Copy Quality Control window opens with only the Display Name category
populated.
3. Edit the categories to create a unique external quality control following the
instructions found in the Adding External Quality Controls section (see Adding
External Quality Controls on page 245).
4. Select Accept to save the edits made to the new external quality control.

Deleting Quality Controls

To delete a quality control:
1. From the appropriate tab on the Quality Control Configuration screen, select the
quality control or manufacturer override to be deleted.
2. Select Delete.
3. Select Yes to confirm the deletion of the selected quality control information.

Loading Unlabeled Quality Controls

When running external quality controls that are not labeled, ensure that no barcodes
are used or that they are not visible through the slots on the Sample Rack during
loading. The operator must categorize the quality control through the Sample
Loading screen during the sample loading process.
To categorize an unlabeled specimen as an EQC during the sample loading
process:
1. Load the EQC into the Sample Bay following the instructions in the Load
Samples section (see Load Samples on page 113). Ensure that the quality
control does not have a recognizable barcode visible through the slots on the
Sample Rack.
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)

Figure 116. Copy Quality Control Window (Example)

System Administration
248