Disable The Lis Connection; Enable Automatic Results Holding - Hologic Panther System Operator's Manual

Table of Contents

Advertisement

on page 117).
To enable LIS settings configured on the LIS Configuration screen:
1. From the Administration screen, select LIS Configuration.
The LIS Configuration screen opens.
2. From the LIS Configuration screen, check the Enable LIS Connection cell.
The Sample Default settings on the Sample Rack Bay screen are disabled. The
system is now configured to run in a bidirectional LIS host query mode. The system
will now interact with the connected LIS to send and receive test order information, as
defined by the options on the LIS Configuration screen.
3. Select Save after altering the LIS configurations.
4. Select OK to confirm.

Disable the LIS Connection

Disable the LIS connection to allow for test order assignment through the Sample
Defaults settings (see Sample Defaults on page 117).
To disable the LIS connection:
1. From the Administration screen, select LIS Configuration.
The LIS Configuration screen opens.
2. From the LIS Configuration screen, uncheck the Enable LIS Connection cell.
The LIS settings on the LIS Configuration screen are disabled. The system will now
assign test orders during Sample Rack loading as defined by the Sample Default
settings on the Sample Rack Bay screen.
3. Select Save after altering LIS configurations.
4. Select OK to confirm.

Enable Automatic Results Holding

If the system is configured to automatically send results to the LIS upon test completion, it
is possible to configure the system to hold results based upon assay and result criteria.
The results that are held are then able to be confirmed, as necessary, and then manually
exported to the LIS through the Results screen (see Sending Results to LIS on page 158).
To enable automatic result holding:
1. From the Administration screen, select LIS Configuration.
The LIS Configuration screen opens.
2. From the LIS Configuration screen, check the Enable Automatic Result Holding
cell.
The Edit Result Types button appears next to the cell.
3. Select Edit Result Types.
The Select Assay to hold from LIS Automatic Sending window opens.
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
System Administration
238