Figure 117. Add Manufacturer Control/Calibrator Override Window (Example) - Hologic Panther System Operator's Manual

Table of Contents

Advertisement

1. From the Manufacturer's Controls/Calibrators tab on the Quality Control
Configuration screen, select Add.
The Add Manufacturer Control/Calibrator Override window opens.

Figure 117. Add Manufacturer Control/Calibrator Override Window (Example)

2. Select the Assay field and select the assay type of the control or calibrator that is
to be updated. Select OK on the selection menu when finished.
3. Select the Ctrl/Cal Name field and select the name of the control or calibrator that
is to be updated. Select OK on the selection menu when finished.
4. Select the desired Result Type and select Edit.
5. Select the Mean field and enter the desired mean value for the control or
calibrator on the keypad. Select OK on the keypad when finished. The mean
must be a value of zero (0) to 100,000,000.
6. Select the Standard Deviation field and enter the desired standard deviation
value for the control or calibrator on the keypad. Select OK on the keypad when
finished. The standard deviation must be a value of zero (0) to 100,000,000.
7. Select the Clear button to remove Mean and Standard Deviation values, if
desired.
8. Select Accept to save the Mean and Standard Deviation values entered for the
chosen manufacturer control or calibrator and return to the Add Manufacturer
Control/Calibrator Override screen.
9. Repeat Steps 4–8 to select Mean and Standard Deviation of additional Result
Types, if applicable.
10. Select Accept to save the override information entered for the chosen
manufacturer control or calibrator and return to the Quality Control Configuration
screen.
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
System Administration
250