Figure 74. Specimens Loaded Panel—With LIS (Example)
Monitoring Control/Calibrator Status
Table 70. Control/Calibrator Status Definitions
Status
Loaded
Valid
Invalid
In-Process
Expired
Pending
Assay Reagents Panel
The control/calibrator set status can also be monitored from the Assay Reagents
panel on the Run Status screen. The Expiration column displays the expiration
countdown time of the control/calibrator set associated with the assay reagent
kit.
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
Meaning
Control/calibrator set has been
loaded.
Processing of the control/calibrator set
is complete and the results are valid.
Control/calibrator set failed due to a
hardware or assay validation failure.
Control/calibrator set is processing.
Control/calibrator results test validity
period has expired.
A control or calibrator assigned a
Pending label will be designated as
white on the rack display.