Table 6. Power Requirements (Continued)
Current Draw
Fuse

Internal Specifications

Table 7. Internal Specifications

Module
Incubators
Incubator 1: High Temperature
(HT)
Incubator 2: Transition (TRAN)
Incubator 3: Amplification (AMP)
Load Stations
HPA
AMP
Bays
Reagent
Sample
Pumps
Vacuum Pump
Ramps
Chiller
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
Average: 700 W
Peak: 1400 W
100 VAC circuit draws 13 amps
l
240 VAC circuit draws 5.4 amps
l
Thermal circuit breaker
Specifications
63.5–64.5 °C
43.2–44.7 °C
42.4–43.0 °C
70.5–73.5 °C
44–46 °C
16.5–19.5 °C
15.5–22.5 °C
4.0–12.0 inHg, target 7.0 inHg
17–19 °C