Table 4. Environmental Requirements (Continued)
Altitude
Ambient Temperature
Relative Humidity
Pollution Degree
Installation Class

Table 5. Environmental Output

Heat Dissipation
Drain Output
Noise Output

Power Requirements

Table 6. Power Requirements

Voltage
Frequency
Current Input
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
Operation: Altitude equal to or less than 2,000 m
(6,562 ft) above sea level
Storage: As required for travel
Operating: 15–30 °C
Storage: 5–45 °C
Transport: -20–70 °C
Operating: 20–85% non-condensing
Storage: 10–90% non-condensing
Transport: 10–90% non-condensing
2
II
550 watts (1878 BTU/hour) during steady state
None: waste fluids drain to removable containers
< 60 db (A)
100–240 ± 10% VAC
50–60 Hz, 1800 VA single phase
Minimum of 15 amp circuit (dedicated)
20 amp circuit (dedicated if used with optional UPS)
Specifications and Requirements