Designating Reflex Tests During Sample Rack Loading; Assay Custom Cutoff Value Configuration - Hologic Panther System Operator's Manual

Table of Contents

Advertisement

2. Check the Enable Reflex Testing cell for the desired reflex condition. The
system will only list the reflex options if both the initial and reflex assays are
loaded onto the system.
The Select at Rack Load option becomes available after checking the Enable
Reflex Testing cell.
3. Check the Select at Rack Load cell.
4. Select Save to save the changes to the configuration setting.

Designating Reflex Tests During Sample Rack Loading

To designate specimens for reflex testing during Sample Rack loading:
1. Load a Sample Rack containing specimens with test orders of the configured
screening assay (assay type found in the Assay column of the Reflex Testing
Configuration screen). If running in batch loading mode, skip Step 2.
In batch loading mode, the screen transitions to the Sample Rack Loading
screen. In random access mode (LIS), specimens that meet the above criteria
are labeled with an "R." Do not close the Sample Bay Door when designating
reflex tests during loading in batch loading mode.
2. If running in random access mode (LIS), select a rack containing specimens with
the "R" notation and then select Rack Details.
3. From the Sample Rack Loading screen, if designating all specimens for reflex
testing, select Reflex All and skip to Step 9. Otherwise, continue to Step 4.
4. If no specimens are to be designated for reflex testing, select Accept Rack.
Otherwise, continue to Step 6.
5. Select the Reflex Assay field to choose the reflex assay for the specimen of
interest.
The Reflex Assay Selection window opens.
6. Select the assay type(s) as desired.
7. Select OK.
8. Repeat Steps 4–7 for any additional specimens to be designated for reflex
testing.
9. When all desired specimens have been designated for reflex testing, select
Accept Rack.

Assay Custom Cutoff Value Configuration

For additional information, refer to the appropriate Panther System Software Application
Sheet.
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
System Administration
253

Advertisement

Table of Contents
loading

Table of Contents