Idle Maintenance; Table 94. Idle Maintenance Procedures - Hologic Panther System Operator's Manual

Table of Contents

Advertisement

will provide an update when there is inadequate volume to perform another Mag Wash
Clean. Perform this task when the instrument is in the Setup or Ready states. Do not add
or remove reagents from the TCR Carousel in the process of performing this task.
Replace the Mag Wash Cleaning Solution as follows:
1. Obtain a new bottle of Mag Wash Cleaning Solution.
2. From the Maintenance Activities screen, select the Replace Mag Wash Cleaning
Solution task.
3. Select Start.
A Run Maintenance Task screen for the Replace Mag Wash Cleaning Solution task
opens.
4. Select Start.
The TCR Door unlocks.
5. Open the TCR Door and the Right Canopy Door.
6. Remove the depleted bottle of cleaning solution.
7. Remove the cap and place the new cleaning solution bottle in the cleaning solution
holder.
8. Dispose of the depleted cleaning solution bottle according to laboratory procedures.
9. Close the Right Canopy Door and the TCR Door.
10. Select Next.
The Maintenance Task Completed screen opens.
11. Select Done.

Idle Maintenance

Store the system as appropriate for any idle duration. If the system is to be left powered on
without processing samples, follow the maintenance activities outlined in the table to
prevent fluid line blockages and deterioration of the subsystems.

Table 94. Idle Maintenance Procedures

Expected Idle
Period
Up to 4 Weeks
Beyond 4 Weeks
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
Procedures
Remove Biohazard Waste prior to expected idle periods
l
Prime the system at least twice per week
l
Continue to perform the Mag Wash Clean procedure as
l
recommended
Contact Hologic Technical Support for additional
l
maintenance information.
Maintenance
210