Figure 79. Results Screen with Filters Panel (Example)

Verifying Results

Once a result has been produced for a test order, it can be verified. After a result has
been verified, no further edits can be made to manually entered or auto-generated
sample IDs. If the Enable Result Verification configuration is enabled, results
cannot be exported to an LIS unless they have been verified (see Enable Result
Verification on page 227).
To verify a worklist:
1. On the Results screen, select the row(s) of the sample result(s).
2. Select Verify.
The system changes the Verified status from "False" to "True." The sample
information can no longer be edited.

Editing Auto-Generated or Manually Entered Sample IDs

If the sample ID was entered either manually (via virtual keyboard or Nonbarcode
Sample Loading) or by using auto sample ID assignment settings, sample IDs can
be changed as long as the sample data has not yet been verified. Follow the
instructions below on how to edit certain sample IDs. (see Enable Auto Sample ID
Assignment for Unreadable Barcodes on page 224).
To edit a sample ID:
1. From the Results screen or the Test Orders screen, select the row of a specimen
with a system-assigned ID.
The Edit Sample ID (Results screen) or Edit (Test Orders screen) button
becomes enabled.
2. Select Edit Sample ID or Edit, as appropriate.
An Edit Test Order confirmation window opens.
3. Select the sample ID field to use the virtual keyboard to enter the sample ID for
the selected specimen. Enter the information on the virtual keyboard and select
OK when finished.
4. Select Accept.
The Results or Test Orders screen updates to reflect the change.
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)
System Operation
155