2. Use the power switch on the rear left of the instrument to turn the instrument hardware
off.
The software recognizes that the computer has been disconnected from the
instrument. Wait until the Instrument Shutdown (Logoff screen) or the Shutdown
(Messages screen) buttons become enabled (approximately 1 minute).
3. From the Logoff screen, select Instrument Shutdown or from the Messages screen,
select Shutdown.
A confirmation window opens.
4. Select Yes to confirm shutdown.
The system software properly shuts down to the Desktop Shield screen.
5. From the Desktop Shield screen, select the shutdown icon.
The computer properly shuts down.
6. To restart the system, follow the System Startup procedure (see System Startup on
page 92).

Replaceable Parts

The instructions in this section contain descriptions of replaceable parts. Always wear
proper protective equipment when replacing Panther System parts.

Vacuum Trap Filter Replacement

l
Universal Fluid Caps Replacement
l
Waste Bottle O-rings Replacement
l
Vacuum Trap Filter Replacement
To replace the Vacuum Trap Filter:
1. Shut down the Panther System (see System Shutdown on page 187).
2. Open the Panther Waste Drawer and locate the Vacuum Trap Filter on the front-right
of the drawer.
®
Panther System
Operator's Manual
AW-17791-001 Rev. 002 (EN)