Error: Tube expected in robot gripper cannot be found
If a tube is expected in the robot gripper during initialization but cannot be found, the
system prompts you to search for the lost tube in the system. When finished, select Start to
continue initialization.
Complete Power Outage in the Laboratory
Problem
AutoMate is not running with an uninterruptible power supply (UPS). The PC needs to be
restarted and tubes are still on the system.
Cause
Complete power outage at the laboratory.
Remedy
1
Turn off the system using the main power supply switch at the rear of the AutoMate.
Ensure that the instrument remains at standstill.
2
Manually remove all tubes from the workplaces and the conveyor belts.
3
Inspect the area in and around the instrument for tubes that might have been dropped
during the power outage. Remove any tubes.
4
Turn on the system using the main power supply switch at the rear of the AutoMate.
5
Press Start
B67471B