Chapter 2 Command System
:CHANnel<n>:OFFSet
Syntax
:CHANnel<n>:OFFSet <offset>
:CHANnel<n>:OFFSet?
Description
Set or query the vertical offset of the specified channel. The default unit is V.
Parameter
Name
<n>
<offset>
Return
The query returns the vertical offset in scientific notation.
Format
Example
:CHANnel1:OFFSet 0.01
:CHANnel1:OFFSet?
:CHANnel<n>:RANGe
Syntax
:CHANnel<n>:RANGe <range>
:CHANnel<n>:RANGe?
Description
Set or query the vertical range of the specified channel. The default unit is V.
Parameter
Name
<n>
<range>
Explanation
This command indirectly modifies the vertical scale of the specified channel (vertical
range=8*vertical scale). The vertical scale can be set by the
command.
Return
The query returns the vertical range in scientific notation.
Format
Example
:CHANnel1:RANGe 8
:CHANnel1:RANGe?
Related
:CHANnel<n>:SCALe
Command
DS1000Z Programming Guide
Type
Range
Discrete
{1|2|3|4}
Related to the current vertical scale and probe
ratio. When the probe ratio is 1,
Real
vertical scale≥500mV/div: -100V to +100V
vertical scale<500mV/div: -2V to +2V
/*Set the vertical offset of CH1 to 10mV*/
/*The query returns 1.000000e-02*/
Type
Range
Discrete
{1|2|3|4}
Related to the current vertical scale and probe
Real
ratio. When the probe ratio is 1,
8mV to 80V
/*Set the vertical range of CH1 to 8V*/
/*The query returns 8.000000e+00*/
RIGOL
Default
--
0V
Default
--
8V