Plumbing connections:
9.
Connect the fluid waste line to port 1 on the HTM valve of the UPLC
heating and trapping module.
10. Connect the fluid inlet line to port 6 on the injection valve of the UPLC
sample manager.
11. Connect the fluid infusion line to the grounded union connected to the
reference valve on the mass spectrometer, which provides lock spray for
the onboard IntelliStart Fluidics.
Requirement:
splitter kit to the fluidics valve before connecting the TRIZAIC fluid
infusion line.
See also:
spectrometer documentation.
Starting up the system
Powering-on the system
On the TRIZAIC source, the LEDs and Stop Flow button are located
Tip:
behind the source door. The LEDs are visible when the door is closed.
To power-on the system:
1.
Power-on the workstation.
Requirement:
other system components.
2.
Power-on the sample manager.
Sample manager
On Xevo TQ instruments, you must fit the supplied flow
nanoACQUITY UPLC Operator's Guide and mass
You must power-on the sample manager before any of the
Fluid waste line
connected to HTM
valve
Fluid inlet line
connected to
Injection valve
Starting up the system
2-9