Technical Data; Operating Device - Waters Andrew Alliance Shaker+ User Manual

Table of Contents

Advertisement

Technical data

Labware
Check the OneLab Domino
compatible
Catalog at
domino-catalog
Mixing frequency
• Tube: 200 to 2000 rpm
• Microplate: 200 to 3000 rpm
(The maximum speed may differ
according to the type of used
labware)
Mixing orbit
2 mm
constant
Speed setting
1 rpm
resolution
Mixing regulation
±25 rpm
accuracy
Relative humidity
80% at 37°C
Maximum
2000 m
operating altitude
Environment
+4 to +37 °C
temperature
operating range
User interface
Through OneLab software
Ingress protection
IP20*
Dimensions
170 x 162 x 100 mm
(W x L x H)
(The specified dimensions are
subject to change according to
the labware)
Weight
2.55 kg
External power
Input 100-240 VAC, 1.4 A / out-
supply
put 24 VDC, 5 A, 120 W
Power or
5 A
current rating
Overvoltage
Cat I
category DC port
Operating voltage
24 VDC
Network
• Gigabit Ethernet
connectivity
• Wi-Fi 2.4 GHz and 5 GHz
Pollution degree
Pollution degree 2
of the intended
environment
Cleaning
Only with a wet tissue or
isopropanol
Warranty
1 year
Certification
CE, FCC, RoHS, WEEE
*Avoid spilling liquid on the device.
andrewalliance.com/
Multiple connected devices
installation
1.
Place your connected
devices as indicated by OneLab.
2.
Open the hatches.
3. Plug the power cables into your devices
power socket.
4. Pass the cables in front of devices and then
insert them into the dedicated path.
5. Close the hatches.

Operating device

Power on
When the device is in standby mode
(plugged to power source but turned
off), press the power button to turn
on the device.
Note: The device will automatically
start up when you plug it to a power
source.
Power off
When the device is in its idle state
5s
(no experiment running), press the
power button for 5 seconds to turn
off the device.