Technical data

Extraction+ pump
Maximum pump rate
Measurement range (absolute)
Measurement accuracy TEB
Resolution
Minimum pressure at 370m
Compatible hose inner diameter
Relative humidity
Maximum operating altitude
Environment temperature operating range
Maximum noise
User interface
Ingress protection
Dimensions (W x L x H)
Weight
External power supply
Power or current rating
Overvoltage category DC port
Operating voltage
Network connectivity
Pollution degree of the intended environment
Cleaning
Warranty
Certification
Extraction+ manifold
Compatible labware
Dimensions (W x L x H)
Weight
Compatible adapters and spacers
*Avoid spilling liquid on the device.
Operating Extraction+ pump
Power on
When the device is in Standby (plug-
ged in to a power source but turned
off), press the power button to turn on the
device.
Note: The device automatically starts
up when you plug it in to a power
source.
13 L/min
0–1030 mbar
25.9 mbar
0.1 mbar
100 mbar
6 mm/1/4 inch
80% at 37 °C
2000 m
+4 to +37 °C
60 dB (A) at 1 m
Through OneLab software
IP20*
200 x 200 x 275 mm
3.7 kg
Input 100–240 VAC, 1.4 A/output 24 VDC, 5 A, 120 W
5 A
Cat I
24 VDC
Gigabit Ethernet/WiFi 2.4 GHz and 5 GHz
Pollution degree 2
Only with a wet tissue or isopropanol
1 year
CE, FCC, RoHS, WEEE
Refer to the OneLab Domino Catalog at
andrewalliance.com/domino-catalog
400 x 127 x 100 mm
1.5 kg
Refer to the OneLab Domino Catalog at
andrewalliance.com/domino-catalog
Power off
5s
When the device is in its idle state (no
ongoing experiment), press the power
button for 5 seconds to turn it off.