3.10 Change of Monitor Graph Display

NOTES: 1.
If measurement is stopped immediately though a
sample remains in the graphite cuvette, the next
measurement will be adversely affected. To
prevent this, the remaining sample should be
eliminated by performing a dummy measurement
without injecting a sample (with a sample cup
removed) before resuming measurement.
2.
Avoid execution of [Maximum Heating] if a sample
remains in the graphite cuvette.
Otherwise, the sample will scatter and
contaminate the inside and vicinity of the cuvette.
Each of monitor graph, profile graph and working curve graph
allows an independent change of its scale and the readout of
cursor-specified data with the trace cursor displayed.
For the above purpose, first bring the mouse cursor onto the
target graph and right-click the mouse to display a popup menu.
(See Fig. 3-52.) In the menu, click the desired item.
Fig. 3-52 Popup Menu
The function of each menu item is explained below.