NOTE:
Only the element after completion of the measurement
dependent on the [Temp. Program Development] can
be reflected on the [Method].
Of the element whose measurement was stopped
halfway or caused an error ("The sample is not
appropriate for the test."), the heating temperature
program will not be generated, so reflection is
impossible.
(2)
Click the [OK] button, and the analytical conditions for the
selected element will be reflected and the [Method] window
will appear.
NOTES: 1.
When reflecting the program onto the [Method],
the [Temp. Program Development] check box is
automatically turned off. For using the Temp.
Program Development function again, turn on the
[Temp. Program Development] check box and
assign the parameters.
2.
If a test is stopped midway and the sequence is
terminated, the [Temp. Program Development]
check box remains turned on in the [Method]
window. For changeover to the usual quantitative
analysis, turn off the [Temp. Program
Development] check box.