3.7 Ignition of Flame

(1)
Close the door at the front of the flame part. Press the
[FLAME ON/OFF] button on the spectrophotometer main
unit for about two seconds.
NOTES: 1.
The [FLAME ON/OFF] button will not function
unless the conditions have been verified.
2.
It is a safety precaution that [FLAME ON/OFF] will
not function unless the button is pressed for about
two seconds.
3.
As the sensor is attached on the front door, be
sure to close the door before igniting.
(2)
A gas leakage test window will now open, so select [Leak
Test (Total System)] and click [OK].
NOTE:
The gas leakage test window opens when the [FLAME
ON/ OFF] button is first pressed on that day.
And the start of measurement will not advance unless
a leakage test is performed.
Make sure to perform a gas leakage test once a day
for safety reasons.
Fig. 3-26 Gas Leakage Test Window
NOTE:
A gas leakage test can be executed as desired by
selecting [Flame Gas Check] on the [Instrument
Control] menu.