10.1 Setting of Temperature Program Test Conditions
NOTE:
When the UNK Replicate is "1", the error (Code: 85-81)
will be displayed for the object element upon clicking
the [Verify] button. (This is because a statistical value
cannot be calculated for "1" entry for this parameter.)
In the [Standards Table] window, 2 or larger value
should be entered for UNK Replicate.
Dry Test
Performs measurement while raising the start temperature
of drying stage at the set intervals to display a graph of
repeatability (RSD) at the drying temperature (Start Temp.).
The end temperature of drying is fixed at the set value.
Put a check mark for [Dry Test] by clicking its check box,
and perform a drying test under the set conditions.
Example Execution of Dry Test
Dry Test Conditions: [Start Temp.], [End Temp.] and
Fig. 10-4 Dry Test Window
10 - 4
Fig. 10-3 Error Message
[Start Temp. Interval] are set at
60°C, 140°C and 20°C.