Required Equipment; Power Supply; Gases - Hitachi ZA3000 Series Instruction Manual

Polarized zeeman atomic absorption spectrophotometer
Table of Contents

Advertisement

1.2 Required Equipment

1.2.1 Power Supply

1.2.2 Gases

The specifications of the power supply for this system are given
in Table 1-1.
Table 1-1 Power Supply Specifications
Part No.
Voltage
7J1-8821
115 V
7J1-8822
220 V
7J1-8823
230 V
7J1-8824
240 V
7J1-8827
220 V
7J1-8828
220 V
Grounding:Connect the instrument to a ground terminal with a
grounding resistance of 100 ohms or less.
Prepare compressed air (compressor), acetylene and argon gases.
The specification of the Argon gas for this system are given
in Table 1-2.
Table 1-2 Argon gas Specification
Gas
Pressure Regulator
Argon
Primary : 25 MPa
Secondary: 1 MPa
Full scale
Attach a stop valve to the terminal of the pipe in the measuring room.
The details refer to the ZA3000 series Maintenance Manual.
NOTE: Operation with an improper gas pressure may result in a
performance degradation of the instrument.
1 - 2
Allowable
Voltage Fluctuation
±10 %
Supply Pressure
(secondary pressure setting)
Capacity
Frequency
50/60 Hz
50 VA
(common)
500 kPa