Hitachi ZA3000 Series Instruction Manual page 44

Polarized zeeman atomic absorption spectrophotometer
Table of Contents

Advertisement

Emission intensity:
Used in emission mode measurement.
 Calculation Mode
Integral is normally used, and need not be changed.
The other modes are mainly used in graphite furnace
analysis.
Integral
: Peak area is converted to area per
Peak Height : Maximum absorbance within the
Peak Area : Area of atomization time – absorbance
Peak Width : Not used in flame analysis. The time
 Wavelength
Normally the wavelength ensuring the highest sensitivity
for the set element is selected as a default value, and
this need not be changed.
When changing to the wavelength of lower sensitivity,
click the [List...] button and the Wavelength dialog box
appear.
Click the [OK] button after selecting the wavelength that
changes.
Fig. 3-4 Wavelength Dialog Box
Moreover, the numerical value can be input directly.
3 - 5
second. This is mainly used in flame
analysis.
atomization time is measured.
is measured.
width when slicing the peak at the
specified height (in %) versus the
maximum absorbance of STD2 is
calculated. In the subsequent
measurements, the slicing height
absorbance to which this time width is
applied is determined, and quantitative
measurement is made based on this
absorbance. The input range of
slicing height is from 1 to 50%.