6.1 Using Correlation Coefficient of Working Curve (Check Working Curve)
6.1.2 Contents to be
printed on the
report
NOTES: 1.
If the correlation coefficient of the regenerated
working curve is still smaller than the reference
value, measurement will be terminated. If the
next analyte element is specified, click the [Next
Element] button to proceed to its measurement.
2.
While Simple STD Add. is selected in
Determination Mode, [Del. Outer STD] cannot be
specified in Action.
The following messages are printed in the report:
 Correlation coefficient is out of range.
 Measurement is finished.
 Delete a standard sample which deviates most from the
working curve and create a working curve again.
 STD□ is deleted. (□ represents an STD number deleted.)
 Re-measure a standard sample which deviates most from the
working curve and create a working curve again.
 Re-measure all the standard samples and create a working
curve again.