Hitachi ZA3000 Series Instruction Manual page 434

Polarized zeeman atomic absorption spectrophotometer
Table of Contents

Advertisement

4.
Shortly after turning on the light source lamp, the
reference signal fluctuates due to a change in the
lamp energy.
The lamp requires a warm-up period of about 10
minutes after turning it on. Measurement should
be started after the reference signal has stabilized.
5.
In a measurement immediately after a cuvette is
mounted, the reference signal may not stabilize
due to contamination of the cuvette.
After mounting a cuvette, carry out heating at the
highest temperature a few times.
The Temp. Program Development function should
be executed after stabilization of the reference
signal.
 Detection Start
Enter the REF absorbance for detection of background
absorption.
Usually, the default value should be used.
Input range: 0.0001 to 1.0000 Abs
 Detection End
Enter the REF absorbance at which it is judged that
background absorption ended after it had been detected.
Usually, the default value should be used.
Input range: 0.0001 to 1.0000 Abs
Dry addition
This function is provided for adding one more drying stage
under the determined conditions after a drying test and then
executing the temperature program.
When you apply check mark to the "Dry addition" check box
by clicking it, stage 2 (drying) will be implemented under the
input conditions.