10.1 Setting of Temperature Program Test Conditions
NOTE:
Whenever the alternate gas is used in [Ash Test], you
must specify [Ash addition] and set a stage for
replacement with the normal gas.
Start Temp. : 550C (or lower)
End Temp. : 550C (or lower)
Ramp Time : 20 s
Gas flow
Gas Type
Atomize Test
Performs measurement while raising the atomization
temperature in the atomization stage at the set intervals to
display a graph of absorption sensitivity (ABS and REF) at
the atomization temperature or that of repeatability (RSD).
Put a check mark for [Atomize Test] by clicking its check
box, and an atomization test can be performed under the
set conditions.
Fig. 10-12 Atomize Test Window
Set the following items.
 Start/End Temp. (C)
Enter the start/end temperature in the [Atomize Test].
Usually enter a temperature a few hundred degrees
lower than the default atomization temperature.
Input range: 50 to 2800C
 Max. Temp. (C)
Enter the maximum temperature when raising the
atomization temperature at the set intervals the [Atomize
Test].
Usually enter a temperature a few hundred degrees
higher than the default atomization temperature or
2800C.
Input range: 50 to 2800C
 Interval (C)
Enter a heating interval for raising the atomization
temperature at the set intervals in the [Atomize Test].
Usually enter a value of about 100C.
Input range: 50 to 500C
10 - 16
: 200 mL/min
: Normal