Hitachi ZA3000 Series Instruction Manual page 325

Polarized zeeman atomic absorption spectrophotometer
Table of Contents

Advertisement

(1)
In Data Table, select the row of the sample to be specified
and click [Sample Blank] - [Set Sample Blank] in the [Data]
menu.
(2)
A dialog box for confirmation opens. Click the
[Recalculate] button.
(3)
To the sample ID set as a sample blank, "S" is attached (for
example, UNK-001S), but the analytical values of that
sample remain unchanged. For the samples after the
sample blank, analytical values after subtracting those of
the sample blank will be indicated.
NOTE:
STD sample cannot be specified as a sample blank.
(4)
For canceling the sample blank specification, select the row
of the corresponding sample and click [Sample Blank] -
[Cancel Sample Blank] in the [Data] menu.
(5)
A dialog box for confirmation opens. Click the
[Recalculate] button.
4 - 27
4.5.11