5.9 Using Monitor Measuring Function

5.9 Using Monitor Measuring Function
ABS values and REF values monitored on the window can be
saved in a designated file as the monitor measurement result
data. For the sample introduction method, manual measurement
is alone available, and the monitor measuremen time range is
from 21 to 3600 seconds (set by the second). Files are saved in
the text format (*.txt). This function is convenient when numeric
data are necessary for a long time.
(1)
Click Start [Data on monitor...] of the [Measurement] menu
(Fig. 5-18).
Fig. 5-18 Measurement Menu Bar
(2)
Specify the destination to save and the file to save with the
folder selection icon
the monitor measurement window (Fig. 5-19). Directly input
the time to monitor in the "Measurement time with
monitoring" or press the right arrow
5 - 22
in the "Monitor Data File" field of
to determine it.