Input of a temperature program requires setting of the
following items.
 Start Temp.
Dry .................... 50 to 1000 C
Ash .................... 50 to 2800 C
Atomize ............. 50 to 2800 C
Clean ................. 50 to 3000 C
NOTE:
The maximum temperature varies with the type of
cuvette: 3000°C for Pyro Tube C HR and Pyro Tube D
HR, 2800°C for Pyro Tube HR and Tube HR, 2600 °C for
platform HR, and 2100 °C for Cup HR. Although entry
is allowed up to 2800 °C on the window, temperature
will not rise beyond the above limits due to the
relationship between cuvette shape and resistance.
 End Temp.
Set the end temperature of dry and ash stages.
You cannot enter each end temperature of atomize and
clean stages because the temperature will be
automatically set at the same temperature as for start.
Dry .................... 50 to 1000 C
Ash .................... 50 to 2800 C
Atomize ............. 50 to 2800 C
Clean ................. 50 to 3000 C
 Ramp Time
Temperature can be raised linearly from the start
temperature to end temperature in the set time.
Input range
Dry .................... 1 to 100 s
Ash .................... 1 to 100 s
3 - 11
(50 to 2800°C for other than Pyro
Tube C HR and D HR)
(50 to 2800°C for other than Pyro
Tube C HR and D HR)
3.1.5