10.1 Setting of Temperature Program Test Conditions
Fig. 10-7 Dry Addition Window
Ash Test
Performs measurement while raising the ashing
temperature in the ashing stage at the set intervals to
display a graph of absorption sensitivity (ABS and REF) at
the ashing temperature.
The start and end temperatures in the ashing stage are
fixed at the same value.
Put a check mark for [Ash Test] by clicking its check box,
and an ashing test can be performed under the set
conditions.
Example Execution of Ash Test
Ash Test Conditions: [Start/End Temp.], [Max. Temp.]
Fig. 10-8 Ashing Test Window
First measurement
Second measurement
Third measurement
Fourth measurement
10 - 10
and [Interval] are set at 400°C,
700°C and 100°C.
Start Temp.
400 °C
500 °C
600 °C
700 °C
End Temp.
400 °C
500 °C
600 °C
700 °C