Measurement Errors; Figure 12: Input Limiting (Clipping) - WAGO 750-495/040-00 Series Manual

3-phase power measurement module 690 v for wago-i/o-system 750 xtr
Table of Contents

Advertisement

56
Function Description
4.4

Measurement Errors

The measurement accuracy of the I/O module is described in section "Device
Description" > "Technical Data" > "Measurement Accuracy." However, the values
indicated only apply if the following conditions are met:
1.
The measurement signals must be within the permitted limits.
At 120 % of the nominal input signals for voltage and current, the
measurement errors increase by 0.10 % for the voltage and 0.15 % for the
current.
The ZC limit (also known as sag level) is
- 30 V
Voltage" setting
- 60 V
Voltage" setting
If the voltage signal is below the ZC limit, no frequency measurement is
possible. The I/O module then outputs the measured value 0 for the
frequency.
A ZC limit underrun is signaled by the "D" LED. If the nominal values of the
I/O module are exceeded, input overloading occurs (see figure).

Figure 12: Input Limiting (Clipping)

In the event of input overloading of the current or voltage measurement
inputs, the I/O module cannot digitize these values and limits the measured
values to the maximum measurable range (clipping of the current or voltage
measured values). As a result, the RMS value that is output is erroneous.
The more the input (and, thus, the reference points for RMS calculation) is
overdriven, the greater the measuring error is. This can also damage the
I/O module.
If the topology is configured as a "3-Wire Wye/Delta", the line-to-line
Manual
Version 1.2.0
750-495/040-00x 3-Phase Power Measurement Module 690 V
if a voltage ≤ 120 V
RMS
if a voltage > 120 V
RMS
WAGO-I/O-SYSTEM 750 XTR
is parameterized for the "Secondary
RMS
is parameterized for the "Secondary
RMS

Advertisement

Table of Contents
loading

This manual is also suitable for:

750-495/040-000750-495/040-001750-495/040-002

Table of Contents